VulnerabilityCVE-2024-9832
No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
Not on CISA KEV
Summary
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the
See the full record, free
- Every source, score and reference in one place
- Timeline and related records
- Watch an entry and get an alert when it changes
Sources and credits
- CVE List Source: CVE Program (cve.org)